NxStage cycler error 23

I was very happy to get an update from my Series One cycler to the VersiHD touch screen cycler.

I love the features of the new machine but it is much more sensitive detecting errors.  I get one or more error 23 – Effluent pressure decreasing on a daily basis.  

I would press Stop followed by Start to correct.  This process added two additional minutes to the run time.

An easier process is to decrease the liters per hour (green triangle) by .1, then immediately increase it by .1.

This resets the pressure limits without stopping and adding an additional two mixtures to the run time.